ANALOG OUTPUT ENABLE/DISABLE SUBMENU:

ANALOG OUTPUT ENABLE/DISABLE SUBMENU:

Press b 5) Scroll though the available selection until ENBL displays to use Analog Output Retransmission (output proportional to the input signal).

Press d 6) Display shows STRD stored message momentarily and then advances to CURR or VoLT Submenu only if it was changed, otherwise press a to advance to CURR or VoLT Current/Voltage Submenu.

If DSBL Analog Output Disabled was selected, all submenus of Analog Output Menu will be skipped and the meter will advance to ALR2 Alarm 2 Menu. If ENBL Analog Output Enabled was selected, Output 1 would be automatically Disabled, and reassigned as Analog Output.

CURRENT/VOLTAGE SUBMENU:

Press

d

7) Display flashes CURR Current or VoLT Voltage.

Press

b

8) Scroll through the available selection: Current or Voltage

 

 

(Example VoLT ).

Press d

9) Display shows STRD stored message momentarily and then

 

 

advances to RD1 Submenu only if it was changed, otherwise

 

 

press a to advance to RD1 Reading 1 Submenu.

READING 1:

 

 

Press

d

10)

Display flashes 1st digit of previous “Reading 1” value.

Press

b & c 11) Enter “Reading 1” value. (Example 0000)

Press d

12)

Display advances to OUT.1 Out 1 Submenu.

OUT 1:

 

 

 

Press

d

13)

Display flashes 1st digit of previous “Out 1” value.

Press

b & c 14)

Enter “Out 1” value. (Example 00.00)

Press d

15)

Display advances to RD!2 Reading 2 Submenu.

READING 2:

 

 

Press

d

16)

Display flashes 1st digit of previous “Reading 2” value.

Press

b & c 17)

Enter “Reading 2” value. (Example 9999)

Press d

18)

Display advances to OUT.2 Out 2 Submenu.

OUT 2:

 

 

 

Press

d

19)

Display flashes 1stt digit of previous “Out 2” value.

Press

b & c 20)

Enter “Out 2” value. (Example 10.00)

Press d

21)

Display advances to the ALR2 Alarm 2 Menu.

The above example is for 0-10 V of the entire range of the Process Input and Analog Output. For 0-20 mA output you need to set “Analog Type” to Current and OUT 2 to 20.00.