3.Align the sensor cable with the swivel mount cable port to prevent cable pinching. Use a 3/8 inch wrench or socket to turn the installation tool shaft clockwise until it is seated in the hole at the top of the sensor flange.

4.Wearing safety face protection, slowly open the isolation valve to the full open position. Loosen the lower hex and jam nuts and move them to the required "H" dimension. Turn the installation tool shaft clockwise until the sensor flange contacts the lower hex and jam nuts. Thread the upper hex nuts down until they contact the sensor flange. Tighten the upper hex nuts to secure the sensor.

5.Remove cap nuts and withdraw the installation tool by turning shaft counterclockwise. Be careful to not damage cable. Snap cable grommet into top of sensor and replace protector plate and cap nuts.

protector plate

installation tool shaft

upper hex nuts

cap nuts

alignment rod

lower hex nuts jam nuts

cap nuts

protector plate

protector plate hex nut

"H"

direction of flow

13