PSF Software Ver.365

6.XON/XOFF: OFF

Please note that these settings should be common with DL-01's.

b-4 Recommended communication settings on DL-01

To configure DL-01 Communication setting, please read "Configuring the software" in the Help topics of DL-01. And select values as follows.

For setting “Type of Device” in the "Settings" panel (Menu—“Edit”--”Settings”),

select “PCS-300/R-100(PCS)/R-300(PSF)” for “R-300 PentaxSpecificationFunction” and other setting should be as follows.

(Please note that these settings should be common with -300's. And if the selection of “Type of Device” is not collect it may result in missing some data.)

R-300BPC(DL-01)

 

Bits per second:

1200 (1200-9600 for sending "SURVEY DATA")

Databits:

8

Parity:

None

Stop bits:

1

PC(DL-01)B-300

 

Bits per second:

1200

Databits:

8

Parity:

None

Stop bits:

1

c) Note for the Memory capacity

Data transfer failure from DL01 to R-300 may cause reduction of memory capacity. If memory capacity becomes less, please back up required data first, and then initialize coordinates data. To initialize coordinates data, turn on the instrument while pressing [F2]+[F5]+[ON/OFF] keys, and take your finger off from [ON/OFF] again.

After you see the message "COORD. DATA INITIAL", press [F5] key. Then the message "Please,

wait" is displayed. When it is completed, the panel of Electronic Vial is displayed.

49

Page 50
Image 50
Pentax R-325(N), R-335(N), R-323(N), R-322(N), R326 instruction manual Recommended communication settings on DL-01