Part 3 Using Support Functions
Hex Code | Explanation | Recommended Action | |
|
|
|
|
64 | The SpO2 module is sending an | 1. | Ensure PicoSat module is |
| error message to the host CPU. |
| properly connected. |
|
| 2. | Replace the PicoSat module. |
|
| 3. | If problem persists, replace main |
|
|
| PCB module. |
|
|
|
|
? | The CO2 module is sending an | 1. | Ensure Oridion module is |
| error message to the host CPU. |
| properly connected. |
|
| 2. | Rplace the Oridion module. |
|
| 3. | If problem persists, replace main |
|
|
| PCB module. |
|
|
|
|
65, 66 | PicoSat SpO2 module has detected | 1. | Replace PicoSat module. |
| an error during initialization. | 2. | If problem persists, replace main |
|
|
| PCB module. |
|
|
|
|
PicoSat SpO2 module has detected | 1. | Replace PicoSat module. | |
| an error on its serial port. | 2. | If problem persists, replace main |
|
|
| PCB module. |
|
|
|
|
Error Code Categories
As a reference, the following table lists the general categories for other error codes. The error code categories are shown only in hexadecimal format.
Hexadecimal Code | Explanation |
|
|
500xxx | internal user interface error |
|
|
501xxx | remove serial port error |
|
|
502xxx | date and time error |
|
|
503xxx | NBP error |
|
|
504xxx | front end error |
|
|
505xxx | alarm error |
|
|
506xxx | audio error |
|
|
507xxx | recorder error |
|
|
508xxx | trend error |
|
|
509xxx | flash memory data error |
|
|
50axxx | SpO2 error |
50bxxx | ECG error |
|
|
50cxxx | |
|
|
50dxxx | |
|
|
50exxx | power monitor error |
|
|
50fxxx | temperature measurement error |
|
|
510xxx | internal user interface error |
|
|
511xxx | error handling error |
|
|
513xxx | serial driver error |
|
|