2 The VitalScreen 17.4" Display
VitalScreen CSR
This display has been designed, tested, and certified for use within the patient vicinity. It has lower electric discharge, thus reducing the likelihood of electric shock. The display is certified to
VitalScreen CR
This display has been designed for hospital use outside the patient vicinity. It is not certified
The VitalScreen CR display has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of FCC rules.
The interface buttons on the VitalScreen CR display are not sealed and can be harmed if exposed to liquids. Use caution when cleaning near the front panel.
Distinguishing the Displays
Many of the features and benefits that differentiate the VitalScreen CSR and VitalScreen CR displays are not visually apparent. The simplest way to identify your Planar display is to check the interface buttons on the front panel.
Note these differences for the interface buttons:
•Interface buttons on the front panel of the VitalScreen CSR display (left) are sealed by a plastic membrane.
•Interface buttons on the front panel of the VitalScreen CR display (right) are not sealed.
The display on the cover of this manual is a VitalScreen CSR. A plastic membrane seals the interface buttons.