TechnicalData
3TechnicalData
3.1MeasurementSpecifications
TemperatureRange
LT‐40to600°C(40to1112°F)
forJThermocouple:‐25to600°C(13to1112°F)
SpectralResponse
LT8to14μm
ResponseTime
Allmodels150ms(95%response)
Accuracy
LT±1%or±1°C2°F)whicheverisgreater
LT±2°C4°F)fortargettemp.<20°C(4°F)
TCoutputs±1%or±2.5°C5°F)whicheverisgreater
Atambienttemperature23°C±5°C(73°F±9°F)
Repeatability
Allmodels±0.5%or±0.5°C1°F)whicheverisgreater