ProgrammingGuide
11.4.3PostProcessing
Thefollowingparameterscanbesettodeterminethepostprocessing
mode,seesection6.4PostProcessingonpage38.
P=5peakhold,holdtime:5s
F=12.5valleyhold,holdtime:12.5s
G=10averaging,averagetime(90%):10s
XY=3advancedpeakhold,hysteresis:3K
XY=2advancedvalleyhold,hysteresis:2K
AdvancedPeak/ValleyHoldwithAveraging:
C=250threshold:250°C
AA=15averagingtime(90%):15s
11.5DynamicData
Alltemperaturerelatedinformationiscalculated128timesasecond.
Torequestthedynamicdata,followingcommandsareavailable:
?Ttargettemperature
?Iinternaltemperatureofthesensinghead
?XJinternaltemperatureofelectronicshousing
?Qenergyvalueoftheinfraredtemperature
?XTtriggersetpoint(active/inactive)fortheFTC3input
Tocheckforresets(e.g.powershutdown)usethecommandXI.
Notice,afteraresettheunitisnewinitialized.
?XIasksfortheresetstatus
!XI0noresetoccurred
!XI1aresetoccurred,newinitializationoftheunit
XI=0setstheresetstatusbackto0
MID71