Installation
5.3.3ThermocoupleOutput
IfyouareusingaJ‐ orK‐ thermocoupleyoumustinstalla
compensationcable.Thecableisavailableasanaccessory
(XXXCI1CB25forTypeJ,XXXCI2CB25forTypeK)withacable
lengthof7.5m(24.6ft.)
Connectthewiresaccordingtothefollowingtable:
J-Thermocouple Power Supply
+
white
red-white
+
red-yellow
yellow
Table1:WiringtheThermocoupleJCompensationCable
K-Thermocouple Power Supply
+
yellow
red-yellow
+
red-white
white
Table2:WiringtheThermocoupleKCompensationCable
Power +
red-
y
ellow
Power
y
ellow
TC J
red-white
TC J +
white
Figure12:WiringtheThermocoupleJCompensationCable
22MID