Installation
5Installation

5.1Positioning

Sensorlocationdependsontheapplication.Beforedecidingona
location,youneedtobeawareoftheambienttemperatureofthe
location,theatmosphericqualityofthelocation,andthepossible
electromagneticinterferenceinthatlocation,accordingtothesections
describedabove.Ifyouplantouseairpurging,youneedtohavean
airconnectionavailable.Wiringandconduitrunsmustbe
considered,includingcomputerwiringandconnections,ifused.
5.1.1DistancetoObject
Thedesiredspotsizeonthetargetwilldeterminethemaximum
measurementdistance.Toavoiderroneousreadingsthetargetspot
sizemustcompletelyfilltheentirefieldofviewofthesensor.
Consequently,thesensormustbepositionedsothefieldofviewis
thesameasorsmallerthanthedesiredtargetsize.Foralist
indicatingtheavailableoptics,seesection3.2OpticalSpecifications
onpage6.
Theactualspotsizeforanydistancecanbecalculatedbyusingthe
followingformula.DividethedistanceDbyyourmodel’sD:S
number.Forexample,foraunitwithD:S=10:1,ifthesensoris
400mm(15.7in.)fromthetarget,divide400by10(15.7by10),which
givesyouatargetspotsizeofapproximately40mm(1.57in.).

14MID