ProgrammingGuide
XFfactorydefaultvalueswillbeset
11.6.5LockMode
Theaccesstotheunitispossibleviaserialinterface(software)and
viathedirectuserinput(modebuttons,LCDdisplay).Itispossible
tolockthebuttons.Thisallowsaccesstheunitonlyviasoftware.
J=Ldirectuserinputviamodebuttonsdenied
Alternativelytheunitcanbeunlockedbypressingthe<Mode/Up>
buttonssimultaneouslyforthreeseconds.
11.6.6ModeSettingfortheDigitalInputFTC3
ThedigitalinputFTC3(seesection5.4.4TriggerandHoldFunction
onpage28)canbeusedasfollows:
XN=TFTC3astrigger
XN=HFTC3withholdfunction
11.6.7ChangingtheSensingHeadCalibrationData
Ifitisnecessarytoexchangethesensinghead,youmustsetthe
calibrationdataforthenewsensinghead:
XZ=01234567FFFFFFFFaccordingtotheheadcalibrationdata
ForMID/MICmodelsthelasteightnumbersarenotusedandmust
besetto„F“likeshownintheexampleabove.ForMIHmodelsand
specials(likeMTBorG5)allsixteennumbersareused.
11.6.8AmbientBackgroundTemperatureCompensation
Incaseofcompensatingtheambientbackgroundtemperature,the
followingmodesareavailable:
AC=0nocompensation
AC=1compensationwithaconstanttemperaturevalueset
withcommandA.
MID73