Basics
4.5ElectricalInterference
Tominimizeelectricalorelectromagneticinterferenceor“noise”be
awareofthefollowing:
Mounttheunitasfarawayaspossiblefrompotentialsources
ofelectricalinterferencesuchasmotorizedequipment
producinglargesteploadchanges.
Useshieldedwireforallinputandoutputconnections.
Makesuretheshieldwiresareearthgroundedatonepoint.
Sensorheadshieldbraidshouldmakedirectcontactaround
thecablecircumference.
MID13