
ProgrammingGuide
11.6DeviceControl11.6.1OutputfortheTargetTemperature
Thesignaloutputcanbesetto4–20mA,0–20mAormV.Ifcurrent
outputisactivated,theoutputcanprovideapredefinedcurrent:
XO=4outputmodeto4–20mA
O=13.57outputofaconstantcurrentat13.57mA
O=60switchesbacktothetemperaturecontrolledoutput
11.6.2AnalogOutput,Scaling
Accordingtothetemperaturerangeofthemodel,itispossibletoset
amaximumvoltage/currentvalueaccordingtoatemperaturevalue
(e.g.,themaximumcurrent20mAshallrepresent200°C/392°F).The
samesettingispossiblefortheminimumvalue.
H=500themaximumcurrent/voltagevalueissetto500°C
L=0theminimumcurrent/voltagevalueissetto0°C
Remark:Youcannotsetthisvalueforthermocoupleoutput.The
minimumspanbetweenthemaximum/minimumsettingsis20K.
11.6.3AlarmOutput
Thesecondoutputchannelcanbesetindifferentmodes,seesection
5.3.2HeadAmbientTemp./AlarmOutputonpage20.
• Internalsensingheadtemperature
• Alarmoutput
K=7internalsensingheadtemperature
K=4alarmoutputforobjecttemperature,0Vincaseofno
alarm
XS=125.3thresholdsettingto125.3°C(ifU=Cisset)
11.6.4Factorydefaultvalues
Itispossibletoresettheunittothedefaultvalues.
72MID