Operation
6.4.5AdvancedValleyHold
Thisfunctionworkssimilartotheadvancedpeakholdfunction,
exceptitwillsearchthesignalforalocalminimum.
6.4.6AdvancedPeakHoldwithAveraging
Theoutputsignaldeliveredbytheadvancedpeakholdfunctions
tendstojumpupanddown.Thisisduetothefact,thatonly
maximumpointsoftheotherwisehomogenoustracewillbeshown.
Theusermaycombinethefunctionalityofthepeakholdfunction
withtheaveragingfunctionbychoosinganaveragetime,thus,
smoothingtheoutputsignalforconvenienttracing.
output temperature
object temperature
without averaging
Temp
Time
Figure29:AdvancedPeakHoldwithAveraging
Theadvancedpeakholdfunctionwithaveragingisonlyadjustable
bymeansoftheDataTempMultiDropSoftware.
6.4.7AdvancedValleyHoldwithAveraging
Thisfunctionworkssimilartotheadvancedpeakholdfunctionwith
averaging,exceptitwillsearchthesignalforalocalminimum.
MID43