Maintenance
9.3SensingHeadExchangeMIHsensingheadsandelectronicboxescannotbe
interchangedwithMID/MICsensingheadsand
electronicboxes!
Theheadexchangerequirestotypeinthenewsensinghead
calibrationdataprintedonthecableasfollows:
1.Toexchangethesensinghead,disconnectthepowerofthe
unit.
2.Connectthewiresforthenewsensingheadaccordingtothe
colordescriptionontheprintedcircuitboard.
3.SwitchthepowerfortheunittoON.
4.Presssimultaneouslythe<Mode/Down/Up>buttons.
5.Fourcharactersappearinthedisplay(formervalues).Typein
thenewdesignator(A)usingthe<Down/Up>buttons.Press
the<Mode>button.
6.Thesecondblockoffourcharactersappearsinthedisplay
(formervalues).Typeinthenewdesignator(B)usingthe
<Down/Up>buttons.Activateyoursettingsbypressingthe
<Mode>button.
B
A
Figure45:SensingHeadCalibrationDataprintedontheCable
(e.g.Headwithtwoblocksof4numbers)
ForMIHmodelsandspeciallymodifiedmodels(likeG5orMTB),
fourblocksoffourcharactersareused.
AlternativelyyoualsocanusetheDataTempMultiDropsoftwarefor
typinginthenewsensingheadcalibrationdata.
MID63