Will clear when:

the low pressure condition is fixed

treatment is stopped by pressing the Start/Stop key.

Power Fail Alarm

Triggered by:

power failure

machine is disconnected or switched off while delivering treatment.

If a power fail alarm sounds:

the flow generator stops delivering air pressure.

Will stop:

when the Alarm Mute key is pressed, or

after two minutes, or

when power is restored.

CAUTION

! Remove the mask from your face if the power fails.

High Pressure Alarm

Triggered by:

the mask pressure exceeds the set trigger level (25 cm H2O) for more than 700 ms.

If a high pressure alarm sounds:

1.The treatment will stop.

2.Turn power off.

3.Check that the air tube and sensor tubes are connected properly.

4.Turn power back on.

5.Remove mask and perform Air Circuit Learn function.

6.Try using the flow generator one more time.

7.If the high pressure alarm activates repeatedly, discontinue use and return to ResMed for servicing. If the alarm does not recur, then continue to use as normal.

Note: The alarm system is designed to ignore coughing. However, if a cough is especially intense and prolonged, it may trigger the high pressure alarm.

Will stop when:

the VPAP Adapt SV is turned off.

High Leak Alarm

Triggered by:

high mask leak (greater than 30 L/min) for more than 20 seconds

the mask being removed while SmartStop is disabled.

If a high leak alarm sounds:

adjust the mask to minimize leak

28

Page 34
Image 34
ResMed Adapt SV System manual Power Fail Alarm, High Pressure Alarm, High Leak Alarm