Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the flow generator.

Problem

Possible cause

Solution

 

 

 

Insufficient air delivered

Air filter is dirty.

Replace air filter.

from flow generator

 

 

 

Air tubing is kinked or

Straighten or replace tubing.

 

punctured.

 

Flow generator operating erratically, ie, switching on and off, pressure varying at random

Water may have entered the air tubing.

Remove the water from the air tubing.

Pressure rises

You are talking, coughing,

inappropriately

voluntarily breath-holding (eg,

 

while rolling over in bed), or

 

intentionally breathing in an

 

unusual manner.

Avoid talking with the mask on, and breathe as normally as possible.

 

Severe mask leak or mouth

Correct your leak condition.

 

leak (greater than 30 L/min).

 

 

 

 

Alarm stays on continuously

Blockage in air circuit.

Ensure there is no barrier to

 

 

the passage of air.

 

Internal malfunction.

Return the device to your

 

 

equipment supplier for

 

 

service.

 

 

 

No LCD display

Power not connected or

Ensure the power cable is

 

switch at back is not on.

connected, and that the

 

 

switch at the back of the

 

 

unit is in the “on” position.

 

 

 

FAULT message displayed

Internal malfunction.

Return the device to your

on LCD

 

equipment supplier for

 

 

service.

System Error 7 message

Device power-on has been

 

interrupted by breathing into

 

the mask during warm up.

Internal malfunction.

Turn the flow generator off and on again. Do not place mask on face during warm up (30 seconds). Select LEARN CIRCUIT (page 22). If the error is not cleared, contact your equipment supplier for service.

Return the device to your equipment supplier for service.

Troubleshooting 35

Page 41
Image 41
ResMed Adapt SV System manual Troubleshooting, Unusual manner