ResMed Adapt SV System manual Technical Specifications

Models: Adapt SV System

1 56
Download 56 pages 57.92 Kb
Page 45
Image 45

Technical Specifications

Dimensions (H x W x D)

5.6" x 9.8" x 11.5" (142 mm x 248 mm x 293 mm)

Weight

Flow generator with cover: 8.14 lb (3.7 kg)

Flow generator with Humidaire 2i (empty): 9.24 lb (4.2 kg)

Environmental conditions

Operating temperature: 41oF to 104oF (+5oC to +40oC)

Storage and transport temperature: -4oF to +140oF (-20oC to +60oC)

Humidity—operating, storage and transportation: 10%–95% non-condensing

Housing construction

Injection moulded plastic

Air tubing

1 x 6’6” (2 m)

Air filter

Synthetic fiber

Patient connection port

22 mm taper, compatible with EN 1281-1:1997 Anaesthetic & Respiratory Equipment —Conical Connectors

Power supply

AC Input range: 110–120 V and 220–240 V; 50–60 Hz; 60 VA

DC Input range (if using DC converter): 12–13.8 V DC

Performance

Operating pressure range: 4–20 cm H2O

Sound pressure level

<30 dBA (tested in accordance with the requirements of ISO 17510-1)

Battery type

Lithium

Alarm specifications

High maximum mask pressure alarm Stops air delivery at 25 cm H2O >700 ms.

High mean mask pressure alarm

Stops air delivery when mean mask pressure exceeds 15 cm H2O (averaged over approximately 1 minute).

Low mask pressure alarm

Mask pressure is less than 3 cm H2O for longer than 5 seconds while the patient is using the VPAP Adapt SV.