Guidance and Manufacturer’s Declaration — Electromagnetic Emissions and Immunity

Guidance and manufacturer’s declaration – electromagnetic emissions

The VPAP Adapt SV is intended for use in the electromagnetic environment specified below. The customer or the user of the VPAP Adapt SV should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment -

guidance

 

 

 

 

 

RF emissions CISPR11

Group 1

The VPAP Adapt SV uses RF energy

 

 

only for its internal function.

 

 

Therefore, its RF emissions are very

 

 

low and are not likely to cause any

 

 

interference in nearby electronic

 

 

equipment.

 

 

 

RF emissions CISPR 11

Class B

The VPAP Adapt SV is suitable for use

 

 

in all establishments, including

 

 

Harmonic Emissions

Class A

domestic establishments and those

IEC 61000-3-2

 

directly connected to the public low-

 

 

voltage network that supplies

 

 

Voltage Fluctuations/Flicker

Complies

buildings used for domestic purposes.

Emissions IEC 61000-3-3

 

 

 

 

 

Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to EMC information provided in this document.

Warnings: The VPAP Adapt SV should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the VPAP Adapt SV should be observed to verify normal operation in the configuration in which it will be used.

The use of accessories (eg Humidifiers) other than those specified in this manual is not recommended. They may result in increased emissions or immunity of the VPAP Adapt SV.

Technical Specifications