SmartStart™

Function

Default

Function Description

Setting Options

 

 

 

 

SmartStart

OFF

Turns the SmartStart function on

ON/OFF

 

 

or off.

 

 

 

See “SmartStart™” on page 23.

 

 

 

 

 

Mask Alarm

ON

Turns the Mask Alarm signal on or

ON/OFF

 

 

off.

 

 

 

If enabled, the Mask Alarm feature

 

 

 

will alert you when a high mask

 

 

 

leak is detected. An audible tone

 

 

 

will sound and a high leak

 

 

 

message will appear on the LCD

 

 

 

screen.

 

SmartStart™

The VPAP III ST-A has a function called SmartStart which can be enabled by your clinician. If SmartStart is enabled, VPAP III ST-A will start automatically when you breathe into the mask and will stop automatically when you take your mask off. This means you do not have to press the Start/Stop key to begin or end treatment.

Note: If you select “Mir Full” as the mask option, SmartStop is automatically disabled. SmartStart may not work with a full face mask due to safety features of the mask.

When Mask Alarm is set to ON, SmartStart/Stop automatically reverts to OFF. SmartStop cannot be used with Mask Alarm because if a high leak occurs, SmartStop will stop treatment before the Mask Alarm signal is activated.

Settings for Mask Types

The following table shows the setting that should be selected for each mask type.

Table 3: Settings for mask types

Settings

Mask

 

 

MIRAGE

Mirage Swift Nasal Pillows System

 

Mirage Nasal Mask

 

 

ULTRA

Ultra Mirage Nasal Mask

 

 

STANDARD

Mirage Activa Nasal Mask

 

Mirage Vista Nasal Mask

 

Modular Nasal Mask

 

 

MIR FULL

Mirage Full Face Mask

 

Mirage Full Face Mask Series 2

 

Ultra Mirage Full Face Mask

 

 

How to Use the VPAP III ST-A Menus

23