ResMed III ST-A user manual Emissions test, Compliance, guidance, Electromagnetic environment

Models: III ST-A

1 56
Download 56 pages 24.16 Kb
Page 49
Image 49
Emissions test

Guidance and Manufacturer’s Declaration - Electromagnetic Emissions and Immunity

Guidance and manufacturer’s declaration – electromagnetic emissions

The VPAP III ST-A is intended for use in the electromagnetic environment specified below. The customer or the user of the VPAP III ST-A should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment -

guidance

 

 

 

 

 

RF emissions CISPR11

Group 1

The VPAP III ST-A uses RF energy only

 

 

for its internal function. Therefore, its

 

 

RF emissions are very low and are not

 

 

likely to cause any interference in

 

 

nearby electronic equipment.

 

 

 

RF emissions CISPR 11

Class B

The VPAP III ST-A is suitable for use in

 

 

all establishments, including domestic

 

 

Harmonic Emissions

Class A

establishments and those directly

IEC 61000-3-2

 

connected to the public low-voltage

 

 

network that supplies buildings used

 

 

Voltage Fluctuations/Flicker

Complies

for domestic purposes.

Emissions IEC 61000-3-3

 

 

 

 

 

Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to EMC information provided in this document.

Warnings: The VPAP III ST-A should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the VPAP III ST-A should be observed to verify normal operation in the configuration in which it will be used.

The use of accessories (eg, humidifiers) other than those specified in this manual is not recommended. They may result in increased emissions or decreased immunity of the

VPAP III ST-A.

System Specifications

41

Page 49
Image 49
ResMed III ST-A user manual Emissions test, Compliance, guidance, Electromagnetic environment