LCD: SYSTEM ERROR

Warning Signal

Cause

Action

 

 

 

The flow generator stops

Component failure.

• Return the VPAP III ST-A for

delivering air pressure.

 

servicing.

Sound: Single

 

• DO NOT USE THE VPAP III ST-A

intermittent tone

 

DEVICE.

LCD: SYSTEM ERROR

 

 

Call service!

 

 

LED: Yellow flashing

 

 

Alarm: System Error

 

 

 

 

 

Sound: Single

High mask leak for

Adjust the mask to minimise leak. See

intermittent tone

more than 20

“Using the Mask-Fitting Feature” on

LCD: HIGH LEAK!!!

seconds.

page 14.

LED: Yellow flashing

 

 

Alarm: Mask off

 

 

Sound: Single

• Air pressure at the

intermittent tone

mask has fallen

LCD: LOW PRES ALARM

below the alarm

LED: Yellow flashing

setting level.

Alarm: Low mask

• Mask is removed

pressure

while SmartStop

 

has been disabled.

1.Check that the air tubing is connected properly.

2.Turn the VPAP III ST-A off and on again at the power switch. If the alarm persists, return the unit to ResMed for servicing.

Sound: Single

Mask pressure

intermittent tone

exceeds alarm setting

LCD: HIGH PRES ALARM

level.

LED: Yellow flashing

 

Alarm: High mask

 

pressure

 

1.The treatment will stop.

2.Turn power off.

3.Turn power back on.

4.Try using the flow generator one more time.

5.If the high pressure alarm activates repeatedly, discontinue use and return to ResMed for servicing. If the alarm does not recur, then continue to use as normal.

The Alarms

31

Page 39
Image 39