LCD: LOW MV ALARM

Warning Signal

Cause

Action

 

 

 

Sound: Single

Minute ventilation

Contact your clinician.

intermittent tone

level has dropped

 

LCD: LOW MV ALARM

below the alarm

 

LED: Yellow flashing

setting level.

 

Alarm: Low minute

 

 

ventilation

 

 

Sound: Single intermittent tone

LCD: NVENT ALARM

LED: Yellow flashing

Alarm: Non-vented mask

Connection of a

• Ensure your mask has an expiratory

 

non-vented mask.

flow port (vent).

Mask expiratory

• Ensure your mask expiratory flow

 

flow port (vent)

ports (vents) are not blocked.

 

may be blocked.

 

Use of

• Contact your clinician.

 

supplemental

 

 

oxygen with a

 

 

vented mask.

 