Warning Signal
The flow generator stops delivering air pressure. Sound: Two alternating continuous tones LCD: LCD turns off LED: Red flashing Alarm: Power fail
Cause | Action | |
• | Power failure. | Alarm will stop: |
• | Power cord is | when the Alarm Mute key is pressed, |
| disconnected or | or |
| VPAP III | after 2 minutes, |
| switched off while | or |
| delivering | when power is restored. |
| treatment (without |
|
| pressing the Start/ |
|
| Stop key). |
|
The flow generator stops | • | Air tubing | 1. | Check that the air tubing is |
delivering air pressure. |
| disconnected from |
| connected properly to the |
Sound: Single |
| the HumidAire 2i/ |
| HumidAire 2i/2iC. |
intermittent tone |
| 2iC. | 2. | Check that the HumidAire 2i/2iC or |
LCD: CHECK TUBE | • | Hardware failure. |
| front cover is connected properly to |
LED: Yellow flashing |
|
|
| the flow generator. |
|
| 3. | Turn the VPAP III | |
Alarm: System fault |
|
| ||
|
|
| again at the power switch. | |
|
|
|
| |
|
|
| If the alarm persists, return the unit to | |
|
|
| ResMed for servicing. | |
|
|
|
| |
The flow generator stops | Treatment pressure | • | Return the VPAP III | |
delivering air pressure. | delivered above a set |
| servicing. | |
Sound: Single | level. | • | DO NOT USE THE VPAP III | |
intermittent tone |
|
|
| DEVICE. |
LCD: PRESSURE ERROR!
LED: Yellow flashing
Alarm: Over pressure
Sound: Single | VPAP III |
intermittent tone | operating outside |
LCD: IPAP LOWER | device specifications. |
ALARM |
|
LED: Yellow flashing |
|
Alarm: Over use |
|
Continue using and contact your clinician about this alarm. Device settings may require adjustment.
The flow generator stops | Hardware error. | • Return the VPAP III |
delivering air pressure. |
| servicing. |
Sound: Single |
| • DO NOT USE THE VPAP III |
intermittent tone |
| DEVICE. |
LCD: SENSOR ERROR! |
|
|
LED: Yellow flashing |
|
|
Alarm: System Error |
|
|
|
|
|
30