ResMed III ST-A Troubleshooting, No display, Insufficient air, Ramp Time is in use, Vpap Iii St-A

Models: III ST-A

1 56
Download 56 pages 24.16 Kb
Page 43
Image 43
Troubleshooting

Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the

VPAP III ST-A.

Problem

Possible Cause

Solution

 

 

 

No display.

Power not connected or

Ensure the power cable is

 

switch at back is not on.

connected and that the

 

 

switch at the back of the

 

 

device is in the ON position.

Insufficient air

Ramp Time is in use.

delivered from the

 

VPAP III ST-A.

 

Wait for air pressure to build up.

 

Air filter is dirty.

Replace air filter.

 

Air tubing is kinked or

Straighten or replace

 

punctured.

tubing.

 

Air tubing not connected

Check air tubing.

 

properly.

 

 

Mask and headgear not

Adjust position of mask and

 

positioned correctly.

headgear.

 

Plug(s) missing from access

Replace plug(s).

 

port(s) on mask.

 

 

Pressure required for

See your clinician to adjust

 

treatment may have

the pressure.

 

changed.

 

 

 

 

The VPAP III ST-A does

Power cord not connected

Connect power cord firmly

not start when you

properly.

at both ends.

breathe into the mask

 

 

(when SmartStart is

Power outlet may be faulty.

Try another power outlet.

enabled).

 

 

 

The VPAP III ST-A is not

Switch power switch at rear

 

switched on.

of the VPAP III ST-A to on (I).

 

SmartStart not on.

Enable SmartStart.

 

Mask Alarm has been

Disable Mask Alarm to

 

enabled; SmartStart has

enable SmartStart if

 

automatically been

appropriate.

 

disabled.

 

Breath is not deep enough to trigger SmartStart.

Take a deep breath in and out through the mask.

Troubleshooting 35

Page 43
Image 43
ResMed III ST-A Troubleshooting, No display, Insufficient air, Ramp Time is in use, delivered from the, Vpap Iii St-A