Getting Started
P5205 Instruction Manual 7
Functional Check
To make a simple functional check of the P5205 probe, select a
source thatsupplies AC line voltage and use the following
procedure.This procedure verifies a majority of the circuitrywithin
the probe. For a complete performance verificati on, refer to the
PerformanceVerification section startingon page 27.
1. Use the installation procedure starting on page 6 t o connect the
output of the P5205 probe to a measurement instrument.
2. Connect the inputs, set the range, and perform the check as each
line of Table1, below, indicates.
Table1: Functional check
Input 1
(+ or --)
Input 2
(-- or+) Mode
Range
setting Check
Hot Ground or
Neutral
Differential 500X
(out)
Measurementinstrument
displaysor indicates the line
voltage
50X
(in)
Overrange indicator lightsif
the input is150 Vp
Hot(same
connection)
Common
Mode
50X or 500X No signal
This completes the functional check proc edure.