Performance Verification
P5205 Instruction Manual 29
Vise
Pliers
ModifiedBNC adapter
Plasticcover removed
Figure 12: Removing plastic covers from binding posts
SetupWARNING. These proceduresrequire the application of high voltage
to the inputs of the P5205 probe. Because t his adapter has exposed
metal surface,only qualified personnel should perform testing with
voltage levels that exceed 30 V rms. All pertinent safety rules and
guidelines for elevated voltage measurements should be followed and
adhered to.
Connect the output of the probe to the Channel 3 input of the test
oscilloscope.Let the probe warm up for 20 minutes.