P5205 Instruction Manual 27
Performance VerificationThe followingprocedure verifies the warranted electrical characteris-
tics of the High Voltage DifferentialProbe. Table 7 itemizes the
equipment required, provides an example or pa rt number of the
equipment, and explains the purpose of the equipment.
Table7: List of equi pment required
Description
Minimum
requirements
Example or part
number Purpose
Testoscilloscop e Bandwidth: ≥400 MHz
TEKPROBE interface
vertical accuracy:
≤1.5%
TDS460A,
TDS744A, or
TDS784A
Displayprobe output
Standard ampli-
tude generator
Leveledsine wave
generator
DC voltage source
Amplitude accuracy:
≤0.75%
Rise time: ≤3ns
Fluke/Wavetek
9100 with oscillo-
scopeoption 250
Check probe attenua-
tion, bandwidth, and
common-mode rejec-
tion ratio
Plunger clamps
(2 required)
Use clampsincluded in
accessory kit
Accessory kit
020-2106-00
Connection to binding
post
ModifiedBNC
adapter
BNC-male-to-dual
bindingpost
103-0035-00 Interconnectionbe -
tween probe and gen-
erator.(See Figure 11
on page 28).
BNC adapter BNC-female-to-dual
banana
103-0090-00 Interconnectionbe -
tween probe and gen-
erator
Coaxial cables
(2 required)
36 in (0.9144 m),
precision50 Ω
012-0482-00 Interconnection
between oscilloscope
and generator
Terminator 50Ωprecision feed
through
011-0049-01 Termination between
probe and generator
during bandwidth mea-
surements