Specifications
16 P5205 Instruction Manual
Table2: Warranted Electrical Characteristics (Cont.)
MaximumOperating Input Voltage 500X differential:±1.3 kV (D C + peak AC)
500X commonmode: ±1kVRMSCAT II
50X differential:±130 V (D C + peak AC)
50X commonmode: ±1kV
RMS CATII
MaximumNondestructive Input Voltage 500X and 50X differential:±1.5 kV (DC +
peak AC)
500X and 50X commonmode: ±1.3 kV RMS
CATII
Gain Accuracy ±3% at 20-30_C, <70%RH
Temperature Operating:0 to 40_C
Nonoperating: -30 to +70_C
Humidity Operating: <85% RH at or below +35_C
Nonoperating: <85% RH at or below
+60_C
Table3: Certifications and Compliances
EC Declaration of
Conformity --Low
Voltage
Compliancew asdemonstrated to the following specification as
listedin the Official Journal of the European Communities:
Low VoltageDirective 73/23/EEC, as amended by 93/68/EEC:
EN 61010-1/A2:1995
Safety requirementsfor electricalequipment for
measurement,control, and laboratory use
EN 61010-2-031:1994
Particularrequirements for hand -heldprobe assemblies for
electricalmeasurement and test equipment
Approvals UL3111 -1 -Standardfor electrical measuring and test equipment
IEC 10106-2-031 -Particularrequirements for hand-held probe
assembliesfor electrical measurement and test
CAN/CSA-C22.2 No. 1010.1-92 and CAN/CSA-C22.2 No.
1010.2.031-94 -Safety requirementsfor electricalequipment for
measurement,control, and laboratory use