Specifications

P5205 Instruction Manual 17

Table3: Certifications and Compliances (Cont.)

InstallationCategory
Descriptions
Terminalson this product may have different installation category
designations.The installation categories are:
CATIII Distribution-levelmains (usually permanently
connected).E quipmentat this level is typically
in a fixedindustrial location
CATII Local-levelmains (wall sockets). Equipment at
thislevel includes appliances, portable tools,
and similar products. Equipment is usually
cord-connected
CATI Secondary(signal level) or battery operated
circuitsof electronic equipment
PollutionDegree 2 Do not operate in environmentswhere conductive pollutants may
be present.
Typical Characteristics

Typicalcharacteristics in Tables 4 and 5 describe typicalbut not

guaranteed performance.

Table4: Typical Electrical Characteristics

Rise Time 3.5 ns
Bandwidth Limit 5MHz
AC Common-Mode Rejection R atio
(20-30_C, <70%RH)
See Figure 5 on page 18
60 Hz: > 10,000:1
100 kHz: > 300:1
100 MHz: > 300:1
AC Noise
(referenced to input)
50X: < 50 mVRMS
500X: < 300 mVRMS
Input Impedance 8M,3.5 pF between inputs
4M,7 pF between each input and ground
SeeFigure6
Propagation Delay 17 nS