Calibrati on-ΤΜ 5006

I NT ERNAL AD JU ST MENT PR OC ED URE

Introduction

This procedure shoul d be performed if t he i nstrument fails to meet the performancei requirements of the elec-

trical characteristics listed n the Specificationi sectio n of this manual . To ensur e continued nstrument accuracy, it is recommended that adjustment be perform ed every 1000 hours of operation or every 6 months if used infrequently. A djustment is also recommended followi ng i nstrument

repair or modificatio n. Adjustments must be made in an ambient temperature of +20°C to +30°C.

Services Available

Tektronix, Inc. pr ovides complete instrument repai r and adjustment at local field service centers and at the factory service center. Contact youri local Tektronix fiel d office or representative fo r further nformation.

Test Equi pment Required

Test equi pment (or equivale nt) listed i n Table 4-1 is required for adjustment of the ΤΜ 5006 . Specificatio ns given for the test equi pment are the minimum necessaky for accurate adjustment. All test equi pment is assume d to be correctly cali brated and operati ng within specificatio n. If ot her testequipment issubstituted, thecalibration setup may need to be alte red to meet t he requirements of the eq ui pment used.

Adjustment Access

R emove the top and lift side covers to gain access to the +8 V buses, ground buses and the adjustments . Fig ure 4-4 shows the adjustment locations.

Power Li mit Adjustment

The Ρωr Lim adjustment, R 2060, is used for troubleshooti ng only . Before beginning calibration make certai n this adjustment is fully cw .

2.Adjust clock Oscillator (SN Β021520 and above)

α. Set the line voltage to 88% of the nominal value fo r the selected range.

b.Connect α frequency counter to pin 1, U 1950 .

ε. ADJU ST- R1855 for α reading of 20 kHz on the counter.

d. Connect α load as described in Fig . 4-3 from +8V to ground.

e.Co nnect α load as described in Fig . 4-3 from +26V to ground.

f.Connect α load as described in Fig . 4-3 from -26V to ground.

All 5.0n

Load for ±26V supplies. Use three 5.091 25W 5% resistors in series (Tektronix P art Number 308-0177-00) .

All 0.3Ω

Load for +8V supply . Use 8 0.3Ω 25W 5% resistors as shown (Tektronix P art Number 308-0691-00) .

2950-45

1 . Adj ust 8 V Adj

α. Connect the dvm to the +8 V and Grid terminalsi as shown in Fig . 4-2. The voltage bus location is shown n Fig. 4-5.

b. Set the line voltage to nominal for the selected range .

c. CHECK -for α reading of 8.2 V .

d. ADJUST- R2151, 8 V Adj fo r α dvm reading of 8.2 V .

Fig . 4-3. Loads for use in internal adjustment procedure.

g.ADJU ST- R1855 slowly εω until the dvm reading just begins to dec rease . Note the frequency reading on the counter.

h. ADJUST- R 1855 εεω for α counter reading 0.25 kHz lower than the reading noted above.

i. R emove all connections.

This completes the internal adjustment proced ure.

REV NOV 1984

4-5

Page 73
Image 73
Tektronix TM 5006 manual Calibrati on-ΤΜ, Services Available, Test Equi pment Required, Adjustment Access, All 0.3Ω

TM 5006 specifications

The Tektronix TM 5006 is a versatile and powerful modular instrumentation platform designed primarily for test and measurement applications. As part of the TM500 series, it provides a wide range of features that cater to various engineering needs, including signal analysis, oscillation measurements, and protocol decoding. It is renowned for its flexible configuration and high-performance capabilities.

One of the standout features of the TM 5006 is its modular design. The chassis can accommodate multiple plug-in modules, allowing users to customize their instrument according to specific requirements. This modularity not only enhances the instrument's usability but also ensures a cost-effective solution, as users can upgrade or change modules as their testing needs evolve.

The TM 5006 supports an array of plug-in modules, including oscilloscopes, waveform generators, frequency counters, and signal analyzers. These modules are engineered to provide accurate and reliable measurements across various parameters, making the TM 5006 suitable for both digital and analog signal testing. Additionally, the ability to integrate different modules facilitates comprehensive testing regimes, enabling engineers to perform a wide spectrum of analyses from a single platform.

Another major technological hallmark of the TM 5006 is its high-resolution display and user-friendly interface. The instrument is equipped with a clear and intuitive interface that allows users to quickly navigate through functions and settings. This ease of use is particularly beneficial in fast-paced testing environments, where time is crucial.

The TM 5006 also boasts advanced measurement capabilities, including high-frequency performance, low-noise operation, and a broad frequency range. Its specifications make it suitable for applications in telecommunications, electronics development, and research laboratories. The built-in triggering mechanisms provide precise control over signal acquisition, while advanced analysis functions enable in-depth signal examination.

In terms of characteristics, the TM 5006 emphasizes reliability and robustness, which are vital for demanding testing environments. The instrument is designed to withstand typical lab conditions, ensuring longevity and consistent performance. Moreover, the TM 5006 complies with various industry standards, providing assurance of its quality and accuracy.

Overall, the Tektronix TM 5006 stands out as an essential tool for engineers and technicians seeking a comprehensive, versatile, and reliable testing platform. Its modular design, advanced measurement capabilities, and user-friendly interface make it an ideal choice for a diverse range of applications in the field of electronics and signal testing.