6712SR Refrigerated Sampler
Section 1 Introduction
1-12
Dry Weight: 120 lbs (54 kg) Refrigerator only, no controller or bottle configuration
135 lbs (61 kg) Includes refrigerator, controller, pump tube, distributor shaft
extension, distributor arm and nut, and discharge tube for the 24 plastic bottle
configuration.
Dry Weight With: 24 plastic bottles, rack, hold down, no caps 150 lbs(68 kg)
24 glass bottles, rack, expander ring, no caps 170 lbs(77 kg)
12 plastic bottles, rack, hold down, no caps 150 lbs(68 kg)
8 plastic bottles, rack, adaptor, no caps 145 lbs(65 kg)
8 glass bottles, rack, adaptor, no caps 155 lbs(70 kg)
2 plastic bottles, rack, no caps 150 lbs(68 kg)
2 glass bottles, rack, no caps 140 lbs(63 kg)
1-9.4 l plastic bottle and cap, no distributor arm or nut 140 lbs(63 kg)
1-9.4 l glass bottle and cap, no distributor arm or nut 145 lbs(65 kg)
1-15.0 l plastic bottle and cap, no distributor arm or nut 140 lbs(63 kg)
Filled Bottle Weight: 24 plastic bottles 101 lbs(46 kg)
24 glass bottles 104 lbs(48 kg)
12 plastic bottles 89 lbs (41 kg)
12 glass bottles 95 lbs (43 kg)
8 plastic bottles 84 lbs (38 kg)
8 glass bottles 90 lbs (41 kg)
4 plastic bottles 98 lbs (45 kg)
4 glass bottles 104 lbs(48 kg)
1 plastic bottle 90 lbs (41 kg)
1 glass bottle 94 lbs (43 kg)
Thermostat Settings: OFF, WARMER, COOLER and 39°F (4°C). The knob turns the
refrigeration system On/Off, and selects the temperature.
Temperature set point
accuracy
±1.8°F ( ±1°C) at 39°F (4°C).
Pull-down time from 75°F
(24°C) to 39°F (4°C)
30 minutes, typical
Recovery time, door open 1
minute with unit operating at
39°F (4°C), 75°F (24°C) ambient
temperature
10 minutes, typical
Voltage 120 Volts, 60 Hz (230 Volts, 50/60 Hz available)
Current Running: 3.3 amp, typical 120V (1.65 amp, typical 230V)
Starting: 12 amp typical 120V (6 amp, typical 230V)
Environment Rain and spray resistant, but not watertight. Do not install in
locations where the refrigerator’s lower compartment could become submerged.
Table 1-3 Technical Specifications for the 6712SR Refrigerated Sampler (Continued)