6712SR Refrigerated Sampler
Section 5 Extended Programming

5-14

4
“LEVEL” SET POINT:
1.000 ft
( 0.001 - 30.00)
Enter a set point of 1 foot.
5
ENABLED WHEN:
ABOVE SET POINT
BELOW SET POINT
Select ABOVE SET POINT.
6
EVENT02:
RAIN LEVEL FLOW
TEMP5 pH5 SP_CO5
DO5 DONE
Select LEVEL again.
7
“LEVEL” CONDITION:
SET POINT
RANGE
RATE OF CHANGE
Choose RATE OF CHANGE.
8
CONDITION IS TRUE
WHEN “LEVEL”
RISES FALLS
Choose RISES.
This will enable the event condition if a rise occurs within the time frame
and above the amount specified in screen #9.
9
“LEVEL” RISES
1.000 ft
__HOURS, __MINUTES
Enter 1 ft and press .
Then enter 2 hours.
10
EVENT03:
RAIN LEVEL FLOW
TEMP5 pH5 SP_CO5
DO5 DONE
Select pH.
11
“pH5” CONDITION:
SET POINT
RANGE
RATE OF CHANGE
Select RANGE.
12
“pHx” RANGE:
LOWER: 4.00
UPPER: 10.00
( 0.00 - 14.00)
Enter 4 for the lower range and 10 for the upper range.
13
ENABLED WHEN:
INSIDE RANGE
OUTSIDE RANGE
Choose INSIDE RANGE.
14
EVENT04:
RAIN LEVEL FLOW
TEMP5 pH5 SP_CO5
DO5 DONE
You will not be entering a fourth event condition, so use the arrow keys to
scroll down, and then select DONE.
15
PACING:
EVENT PACED
3 EVENT CONDITIONS
Proceed with the rest of the programming steps.
Table 5-3 Event Paced Sampling (Continued)