6712SR Refrigerated Sampler
Section 5 Extended Programming
5-9
35
ENABLE:
RAIN LEVEL FLOW
NONE
Select RAIN.
Note: All valid enable options will appear here, so your screen may differ.
36
ENABLE: RAIN
AND OR DONE
Select AND.
37
ENABLE: RAIN AND
LEVEL FLOW NONE Select LEVEL.
38
“RAIN” SET POINT:
0.15 INCHES PER
__HOURS, __MINUTES
Enter 0.15 for inches, then enter 30 minutes.
39
ENABLED WHEN:
ABOVE SET POINT
BELOW SET POINT
Select ABOVE SET POINT.
40 RESET RAIN HISTORY?
YES NO
Select YES.
41
“LEVEL” CONDITION:
SET POINT
RANGE
RATE OF CHANGE
Select SET POINT.
42
“LEVEL” SET POINT
0.25 ft
(0.001-30.000)
Enter 0.25.
43
ENABLED WHEN:
ABOVE SET POINT
BELOW SET POINT
Select ABOVE SET POINT.
44
ONCE ENABLED,
STAY ENABLED?
YES NO
Select YES.
45 SAMPLE AT ENABLE?
YES NO
Select YES.
46
PAUSE RESUME
1.HH:MM DD HH:MM DD
2.HH:MM DD HH:MM DD
CLEAR DONE
Select DONE.
Screen will say “End of Part A” then “Beginning Part B.”
47
UNIFORM TIME PACED
FLOW PACED
EVENT PACED
NONUNIFORM TIME
Select FLOW PACED.
Table 5-2 Storm Water Runoff Sampling (Continued)