6712SR Refrigerated Sampler
Section 5 Extended Programming
5-18Figure 5-1 Sample Distribution

Sequential

123456

Bottles-per-Sample

112233

Samples-per-Bottle

1357
2468

Multiple Bottle Compositing

114477
225588
3366

Time Switched (Flow-paced, No sample at start)

1
2
TO
4
5
6
7
8
EXTENDEDPROGRAMMINGONLY
DISTRIBUTION:
1 BOTTLES/SAMPLE
SWITCH AT 13:00
THEN EVERY 1:00
1 BOTTLES PER
SAMPLE EVENT (1-24)
SWITCH BOTTLES ON:
NUMBER OF SAMPLES
TIME
SWITCH BOTTLES EVERY
1 HOURS, 0 MINUTES
FIRST SWITCH TIME
AT: 13:00
Start Time TimeOverride
EXTENDED
PROGRAMMING
DISTRIBUTION:
SEQUENTIAL 1 BOTTLES PER
SAMPLE EVENT (1-24)
SWITCH BOTTLES ON:
NUMBER OF SAMPLES
TIME
SWITCH BOTTLES EVERY
1 SAMPLES (1-50)
YES NO
RUN CONTINUOUSLY?

Each bottle receives one sample from one sample event.

EXTENDED
PROGRAMMING
DISTRIBUTION:
2 BOTTLES/SAMPLE 2 BOTTLES PER
SAMPLES EVENT (1-24)
SWITCH BOTTLES ON:
NUMBER OF SAMPLES
TIME
SWITCH SETS EVERY
1 SAMPLES (1-50)
YES NO
RUN CONTINUOUSLY?

Multiple bottles receive a sample from one sample event.

EXTENDED
PROGRAMMING

Each bottle receives a sample from multiple sample events.

DISTRIBUTION:
2 SAMPLES/BOTTLE 1 BOTTLES PER
SAMPLE EVENT (1-24)
SWITCH BOTTLES ON:
NUMBER OF SAMPLES
TIME
SWITCH BOTTLES EVERY
___ SAMPLES (1-50)
YES NO
RUN CONTINUOUSLY?
EXTENDED
PROGRAMMING
ONLY
DISTRIBUTION:
2 BOTTLES/SAMPLE
3 SAMPLES/BOTTLE
2 BOTTLES PER
SAMPLE EVENT (1-24)
SWITCH BOTTLES ON:
NUMBER OF SAMPLES
TIME
SWITCH BOTTLES EVERY
3 SAMPLES (1-50)
YES NO
RUN CONTINUOUSLY?