6712SR Refrigerated Sampler
Section 5 Extended Programming

5-24

5.12.1 Sampler Enable This example uses a sampler with an attached 730 Bubbler Flow
Module, Rain Gauge and an SDI-12 sonde at address 5 with four param-
eters.
Table 5-4 Sampler Enable
1
ENABLE:
NONE PROGRAMMED
In this example, you will be enabling the sampler if it rains .5” in 2 hours
and the level goes above 1 foot.
2
ENABLE:
RAIN LEVEL FLOW
TEMP5 pH5 SP_CO5
DO5 NONE
Select RAIN.
3
ENABLE: RAIN
AND OR DONE
Choose AND.
4
ENABLE: RAIN AND
LEVEL FLOW TEMP5
pH5 SP_CO5
DO5 DONE
Select LEVEL.
5
“RAIN” SET POINT:
.5 INCHES PER
__HOURS, __MINUTES
Enter .5 for the inches of rain. Press
.
Enter 2 HOURS.
6
ENABLED WHEN:
ABOVE SET POINT
BELOW SET POINT
Select ABOVE SET POINT.
7RESET RAIN HISTORY?
YES NO
Select YES.
8
“LEVEL” CONDITION
SET POINT
RANGE
RATE OF CHANGE
Select SET POINT.
9
“LEVEL” SET POINT:
+1 ft
( 0.001 - 30.00)
Enter 1 foot.
10
ENABLED WHEN:
ABOVE SET POINT
BELOW SET POINT
Select ABOVE SET POINT.
11
ENABLE:
RAIN >0.50” / 2:00
AND
LEVEL >1.000 ft
Your enable conditions will be summarized on the display.
If correct, press to continue.
If necessary, make changes to the enable conditions.