6712SR Refrigerated Sampler
Section 5 Extended Programming
5-10
48
PACED BY:
FLOW PULSES
FLOW MODULE VOLUME
Select FLOW MODULE VOLUME.
49
FLOW BETWEEN
SAMPLE EVENTS:
2500 gal
(0.001-99999)
Enter 2500.
50 SAMPLE AT START?
YES NO
Select YES. This will cause the first sample to be taken 20 minutes (the
delay to start of sampling specified in Step 67) after the enable for Part B.
51 1 BOTTLES PER
SAMPLE EVENT (1-18)
Enter 1.
52
SWITCH BOTTLES ON:
NUMBER OF SAMPLES
TIME
Select NUMBER OF SAMPLES.
53 SWITCH BOTTLES EVERY
1 SAMPLES (1- 50)
Enter 1.
54
RUN CONTINUOUSLY?
YES NO
Select NO.
55 SAMPLE VOLUME:
200 ml (10-1000)
Enter 500.
56
ENABLE:
RAIN LEVEL FLOW
‘A’ DONE NONE
Select RAIN.
57
ENABLE: RAIN
AND OR DONE
Select AND.
58
ENABLE: RAIN AND
LEVEL FLOW ‘A’ DONE
DONE
Select LEVEL.
59
“RAIN” SET POINT:
0.15 INCHES PER
__HOURS, __MINUTES
Enter 0.15 for inches, then enter 30 minutes.
60
ENABLED WHEN:
ABOVE SET POINT
BELOW SET POINT
Select ABOVE SET POINT.
Table 5-2 Storm Water Runoff Sampling (Continued)