6712SR Refrigerated Sampler
Section 4 Standard Programming
4-31
8
MODE OF OPERATION:
FLOWMETER
LEVEL ONLY
If you choose Flowmeter, proceed to screen #9 below.
If you choose LEVEL ONLY, proceed to screen #12.
9
WEIR FLUME
DATA POINTS
EQUATION MANNING
FLOW-INSERT
When you select Flowmeter, you will have several options for performing
the flow conversion.
For this example, choose Weir. Press .
10
V-NOTCH
RECTANGULAR
CIPOLLETTI
For this example, choose V-Notch. Press .
11
V-NOTCH WEIR ANGLE:
120 90 60
45 30 22.5
For this example, choose a Weir angle of 90. Press .
12
CURRENT LEVEL IS
__.__ ft
ADJUST LEVEL TO
__.__ ft
The current level will be displayed. Enter an adjustment if needed. If the
adjustment is different by more than 0.50 ft, you will be prompted “ARE YOU
SURE? YES NO” Choose NO to re-enter; select YES if correct. Press .
13 NUMBER OF BOTTLES:
12481224
Select the number of bottles in your bottle kit by pressing either arrow key
until the correct number blinks. Press . For this example, select 24.
14 BOTTLE VOLUME IS
1000 ml (300-30000)
Type the volume for the bottles in your kit. For this example, 1000 is
correct, so simply press .
15 SUCTION LINE LENGTH
IS 10 ft
(3-99)
Type the length of the suction line, then press .
If you change the length, the sampler will display a message,
“PLEASE WAIT!. . . GENERATING PUMP TABLES.”
16 TIME PACED
FLOW PACED
Select FLOW PACED.
If the module is being used as a flowmeter, you will see screen #17a.
If the module is being used for level only, you will see screen #17b.
17
a
FLOW BETWEEN
SAMPLE EVENTS:
0.075 Mgal
(0.001 - 99999)
If the module is being used as a flowmeter, choose the amount of flow
between sample events.
Table 4-4 Programming Example With 730 Module Installed (Continued)