6712SR Refrigerated Sampler
Section 4 Standard Programming
4-4
4NUMBER OF BOTTLES:
12481224
Select 24.
5BOTTLE VOLUME IS
1000 ml (300-30000)
Enter 1000.
6SUCTION LINE LENGTH
IS 10 ft
(3-99)
Enter 7.
7TIME PACED
FLOW PACED
Select FLOW PACED.
8
FLOW BETWEEN
SAMPLE EVENTS:
2 PULSES (1-9999)
Enter the number of pulses between sample events.
9
SEQUENTIAL
BOTTLES/SAMPLE
SAMPLES/BOTTLE
Select BOTTLES/SAMPLE.
10
RUN CONTINUOUSLY?
YES NO
For this example, select NO by pressing an arrow until the option blinks.
Then, press . Selecting YES allows the program to run indefinitely by
repeating the sample distribution. Continuous sampling assumes that filled
bottles are replaced with empty bottles at regular service intervals.
11 1 BOTTLES PER
SAMPLE EVENT (1-24)
Enter 2.
12
NO DELAY TO START
DELAYED START
CLOCK TIME
WAIT FOR PHONE CALL
Select CLOCK TIME.
13
START FLOW COUNT AT:
00:00 Enter 6, then enter 0.
This is the time at which the sampler will start the pacing countdown.
14
SELECT DAYS:
SU MO TU WE TH FR SA
DONE
Select MO, WE, and FR.
Use the arrow keys and to select the days. When the correct days are
blinking, select DONE and press ↵.
15
PROGRAMMING COMPLETE
RUN THIS PROGRAM
NOW?
YES NO
Select NO.
Table 4-2 Standard Program: Flow-Paced Sampling, Two Bottles Per Sample (Continued)