Flue Gas Analysis System

General Information

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TABLE 1: Detector Response To GASES

COMPOUND

LEL*

RESPONSE FACTOR

Methane

5.0

1.00

Hydrogen

4.0

0.86

Carbon Monoxide

12.5

0.32

Ethane

3.0

1.20

Ethylene

2.7

1.26

Acetylene

0.5

1.39

Propane

2.2

1.42

Propylene

2.0

1.33

Butane

1.9

1.54

Hexane

1.1

1.50

Cyclohexane

1.3

1.44

Heptane

1.05

1.59

Benzene

1.3

1.50

Pentane

1.5

1.45

Toluene

1.2

1.48

Ethylene oxide

3.6

0.76

Methyl Ethyl

1.8

0.96

Ketone

 

 

Methyl Acrylate

2.8

0.59

*Taken from Fire Hazard Properties of Flammable Liquids, Gases and Volatile Solids, National Fire Protection Agency.