|
| Flue Gas Analysis System | Operation 4 | |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
4.2.2Span Gas
It is recommended that the span gas have a concentration of
Note: It is important that the dilution air flow is never turned off during operation or calibration. Combustibles without at least a 2 : 1 ratio of oxygen should never be allowed to flow past the sensor. If this occurs, a loss of sensitivity will result and the unit will require complete recalibration and possible sensor replacement.
Note: The range of the combustible monitor is
4. 3 System Startup
4.3.1Analyzer Startup
Refer to Sections 4.1 and 4.2 for startup procedures for the oxygen and combustibles analyzers, respectively.
4.3.2Sample System Startup
Start the pump by turning ON the oxygen analyzer RANGE switch. Then turn on the water by opening the shutoff valve inside the case. Water should be observed passing through the transparent plastic hoses. Water flow rate will be determined by the supply pressure. With 10 psig, flow rate will be approximately 1/4 gpm maximum pressure should be 100 psig. This will give a flow rate of about 11/2 gpm. Generally, the dirtier the sample, the more water required. However, a 1/4 gpm flow rate will nor- mally be adequate, and the pressure should be reduced accordingly.
After analyzer startup and calibration procedures have been accom- plished, make certain that the
TELEDYNE BROWN ENGINEERING |
Analytical Instruments