General Information

Model 9700

 

 

 

 

3 Oxygen Analyzer Section

The analysis system (see Figure 3) employs a Model 326B Analyzer which features a TBE/AI unique, patented Micro-Fuel Cell oxygen sensor. This electrochemical transducer provides an electrical signal that is directly proportional (and specific) to the oxygen concentration in the gas phase immediately adjacent to its sensing surface.

The analyzer is capable of oxygen measurements as small as 0.05% and is unaffected by flow rate changes. Since the sensor is incapable of producing a significant electrical signal without the presence of oxygen, any measurable signal produced is “specific” to oxygen.

The Micro-Fuel Cell is completely enclosed, maintenance-free, and has a predictable life span that is covered by warranty. When the cell is expended, it is discarded in the same fashion as a worn-out flashlight battery. TBE/AI’s extensive line of Micro-Fuel Cell equipped oxygen- measuring instruments are all designed so that the sensor cell can be replaced in a matter of moments by nontechnical personnel without the use of tools.

3 .1 Standard Features

The oxygen analyzer is housed in a fiberglass equipment case de- signed to resist the invasion of moisture and dust. The following features are standard in the Model 326B analyzer:

Three Ranges of Analysis

Three ranges of analysis are offered 0–5% (Lo), 0–10% (MED), and 0–25% (HI). Range control is achieved through the positioning of a panel-mounted RANGE selector switch; the switch also controls the power on/off to the analyzer. Since one

of the three available ranges of analysis is always 0–25% O2. ambient atmospheric air (20.9% O2) can always be used to calibrate the analyzer. A CAL mark on the oxygen meter scale facilitates calibration. (Although the ranges selected best cover the oxygen content of flue gas, any three ranges of analysis, from 0–1% to 0–100%, can be provided. )

4

TELEDYNE BROWN ENGINEERING

Analytical Instruments

Page 8
Image 8
Teledyne 9700 instruction manual Oxygen Analyzer Section, Standard Features