General Information | Model 9700 |
|
|
|
|
The sample is sent through one side (SAMP) of twin indicating flowmeters. Since it is necessary that there is sufficient oxygen in the sample being analyzed to insure full combustion of any combustible gases present, the sample is blended, or diluted, with an equal amount of “clean” compressed air. The compressed air is introduced through the second (AIR) of the two flowmeters. Two flowmeters are used so that an equal volume of both sample and air flow can be visually set.
The flowmeters, valves, sensor and associated plumbing are installed within a temperature controlled box. Temperature is held at about 130 deg F to keep all components above the dew point of the sample gas.
The sample stream is routed past the combustible gas sensing element. This element is a
The beads are installed in a housing which has a flashback arrestor screen at the sensing aperture to prevent flame propagation back into the process.
Response of a catalytic bead sensor to a number of gases is shown in Table I.
An adjustable alarm can be set at any value within the full range. The latching or
At completion of combustibles analysis, the sample stream is vented from the analysis circuit.
10 | TELEDYNE BROWN ENGINEERING |
Analytical Instruments