5
Flue Gas Analysis System General Information
TELEDYNE BROWN ENGINEERING
Analytical Instruments
Integral Meter Readout
The analyzer is equipped with a 5" panel meter for direct
readout of the analysis. A linear scale (mirror equipped to
eliminate parallax) promotes reliable, accurate readout of the
analysis at any point on the scale. Resolution and accuracy of
the meter eliminates the necessity of an accessory readout
device, unless permanent recording or remote indication is
required.
Output Signal
A linear output signal of from 0–1mV dc to 0–1 V dc is
available for those applications requiring a remote indication
and/or recording of the sample oxygen. Unless otherwise
specified, the output signal will be 0–1 V dc.
Temperature Control and Compensation
To eliminate the inaccuracies caused by varying temperature
conditions that are inherent in most methods of analysis em-
ploying transducers, a system composed of a combination of
temperature compensation and control is used in the analyzer.
Figure 3: Oxygen Analyzer Section