Flue Gas Analysis System

General Information

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 3: Oxygen Analyzer Section

Integral Meter Readout

The analyzer is equipped with a 5" panel meter for direct readout of the analysis. A linear scale (mirror equipped to eliminate parallax) promotes reliable, accurate readout of the analysis at any point on the scale. Resolution and accuracy of the meter eliminates the necessity of an accessory readout device, unless permanent recording or remote indication is required.

Output Signal

A linear output signal of from 0–1mV dc to 0–1 V dc is available for those applications requiring a remote indication and/or recording of the sample oxygen. Unless otherwise specified, the output signal will be 0–1 V dc.

Temperature Control and Compensation

To eliminate the inaccuracies caused by varying temperature conditions that are inherent in most methods of analysis em- ploying transducers, a system composed of a combination of temperature compensation and control is used in the analyzer.

TELEDYNE BROWN ENGINEERING

5

Analytical Instruments

Page 9
Image 9
Teledyne 9700 instruction manual Oxygen Analyzer Section