5 Maintenance & Troubleshooting

Model 9700

 

 

 

 

Note: Do not over-order or stockpile Micro-Fuel Cells. Only one cell per instrument should be kept in reserve.

When installing the oxygen cell in the probe housing, make certain that the shorting clip is re- moved and that the cell membrane is facing up (or outward). Do not install the oxygen cell upside down.

Caution: When replacing the oxygen cell, use care not to scratch the membrane covering the gold-plated electrode. If the membrane is ruptured or damaged in any way, the sensor must be re- placed.

The oxygen cell contains a caustic liquid (potassium hydroxide solu- tion). If the Teflon membrane is ruptured, the liquid in the cell can leak out. The liquid has a characteristic slippery feel and if contact with the skin or eyes occurs, wash with copious amounts of water and seek medical attention, if required.

Note: Do not install an oxygen cell into a system that is in storage or in a non-operating condition. The oxygen cell will commence the oxida- tion process and eventually will become useless as a sensor. Also, when the on-line system is shut down for a short time, every effort should be made to adjust the controls in the system to prevent air from coming into contact with the Micro-Fuel Cell.

5.1.3Cell Warranty

The A-3 Micro-Fuel Cell carries a warranty that covers its normal life expectancy. Cell warranty is for 6 months of continuous service in normal flue gas applications.

Customers having warranty claims should return the cell in question to the factory for evaluation. If it is determined that failure is due to faulty material or workmanship, the cell will be replaced free of charge.

Note: Evidence of tampering or abuse will render the warranty null and void. If the cell was working satisfactorily, but fails short of its warranty period, the customer will receive credit, on a prorated basis, towards the purchase of a replacement cell.

5-2

TELEDYNE BROWN ENGINEERING

Analytical Instruments

Page 38
Image 38
Teledyne 9700 instruction manual Cell Warranty