Flue Gas Analysis System
Maintenance & Troubleshooting 5
Maintenance & Troubleshooting
5.1Maintenance
5.1.1 Routine Maintenance
No moving parts other than the meter movement and the relay con- tacts are contained in the analyzers. Periodic service, therefore, other than oxygen cell replacement, is not required. The periodic calibration con- tained in Sections 4.1 and 4.2 should be adequate to keep the analyzers functioning. If troubles develop, refer to the trouble shooting procedures contained in Section 6.
5.1.2Oxygen Cell Replacement
When the
Wipe off contact pins (P/N
To offset the possibility of not having a replacement cell available when it is needed, TBE/AI recommends that a spare cell be ordered shortly after the instrument is placed in service, and each time the cell is replaced thereafter.
TELEDYNE BROWN ENGINEERING |
Analytical Instruments