Model OT-2 System Operation 3

3-9

Teledyne Analytical Instruments

3.3.3 H2S Scrubber

The life expectancy of the stainless steel H2S scrubber, in-
cluded in this system, is determined by the amount of H2S passing
through it.
With the maximum concentration of 200 PPM H2S and a flow
rate of 0.2 scfh the expected scrubber life will be over 6 months
with Purifil® and over 11 months with Purifil® II.
This scrubber life expectancy increases with lower H2S con-
centration or gas sample flow rates. It is recommended that the
flow rate be set no higher than 0.4 scfh initially. The flow rate
should be increased only if a faster system response time is de-
sired.
It is suggested that a spare scrubber be kept on hand and
replaced every 6 months. When the scrubber needs to be re-
placed, follow the procedure in section 4,

Maintenance

, "Scrub-
ber Replacement".
WARNING: IF SCRUBBER IS NOT PROPERLY MAINTAINED THE
O2SENSOR LIFE WILL BE SHORTENED.

3.3.4 Exhaust/Vent Line (Sample Return)

It is recommended that a long vent line (approx. 4 ft.) be
installed to the "Sample Return" connector at the side of the sys-
tem enclosure. The extended vent line will minimize back diffusion
from the atmosphere back to the sensor. The minimum recom-
mended tubing diameter is 1/4" to minimize back pressure.
3.4 Testing the System
Before connecting the instrument into the power source:
Check the integrity and accuracy of the gas
connections. Make sure there are no leaks.
Check the integrity and accuracy of the electrical
connections. Make sure there are no exposed
conductors.
Check that inlet sample pressure is within the
acceptable range (see section 3.3.1).
Power up the system, and test it.