Model OT-2 System

Maintenance 4

 

 

 

 

 

 

Maintenance

4.1Routine Maintenance

Aside from normal cleaning and checking for leaks at the gas connections, routine maintenance includes servicing the filter, the scrubber, installing replacement Micro-Fuel cells and recalibration. For recalibration, see section 3.3 Calibration.

4.2Cell Replacement

The Micro-Fuel Cell is a sealed electrochemical transducer with no electrolyte to change or electrodes to clean. When the cell reaches the end of its useful life, it is replaced. The spent fuel cell should be discarded according to local regulations.

The characteristics of the Micro-Fuel Cell are similar to those of a mercury battery in that both provide an almost constant output (when the sensor is exposed to a constant level of oxygen) throughout their useful life, and then fall off sharply towards zero at the end. Cell failure, in the Model OT-2 System, will probably be indicated by the inability to properly span calibrate the transmitter, or excessively long response time to changes in oxygen concen- tration.

To ensure availability, TAI recommends that a spare cell be purchased shortly after the instrument is placed in service, and each time the cell is replaced.

The spare cell should be carefully stored in a cool, and near constant temperature, area if possible.

NOTE: Do not disturb the integrity of the cell package until the cell is to actually be used. If the cell package is punctured

and has been exposed to atmospheric air over an ex- tended period, the sensor may not function properly.

Power must be removed from the OT-2 prior to removing or installing the sensor.

Teledyne Analytical Instruments

4-1

Page 29
Image 29
Teledyne operating instructions Model OT-2 System Maintenance