2 Operational Theory | Model | |
|
|
|
|
|
|
2.6Sample System
The system consists of a shut off valve, particulate filter, pres- sure regulator, scrubber, calibration selector valve, flow control valve, O2 transmitter and a sample flow meter and incorporates ¼ inch tube fittings for sample inlet and outlet connections at the side of the enclosure.
H2S from the sample gas can diffuse into the
Sample Flow Description: (See Figure 2-4)
1.The sample enters the transmitter system via the shut off valve at
2.A pressure regulator, which has been preset at the factory to approximately 5 psig ensures a constant gas flow to the sensor.
3.After the pressure regulator, the sample gas passes through the scrubber to the calibration selector valve.
4.When the sample has been selected, it flows through the flow control valve to the O2 sensor.
5.Finally, the sample flows through the flow meter and exits the system through the vent line to a safe area at atmospheric pressure.
Note: The flow control valve should be adjusted until the flow meter indicates a flow rate of 0.4 scfh.
|
| H2S | FLOWMETER |
|
SAMPLE IN | FILTER | SCRUBBER |
| |
|
|
| VENT TO | |
|
|
| O2 XMTR | ATMOSPHERE |
|
|
| PRESSURE | |
|
|
| (SAFE AREA) | |
|
|
| SPAN GAS IN |
|
Figure 2-4 Flow Diagram
Teledyne Analytical Instruments |