4 Maintenance4 Maintenance4 Maintenance4 Maintenance4 Maintenance Model OT-2 System
To replace the cell: Figure 4-1
1. Open the enclosure door.
2. Undo the velcro straps.
3. Remove the insulation.
4. The cell block is located in the center of the unit.
Unscrew the cap on the bottom of the cell block. The
B-2C sensor should drop down once the cap is
removed.
Figure 4-1 Cell Block
NOTE: Before the cell is removed from its package, a zero gas
with low ppm oxygen flowing at a relative high rate
through the cell block can lessen the time it will take the
newly installed cell to drop to a zero reading.
3. Remove the new cell from its package, and carefully
remove the shorting clip.
WARNING: The sensor used in the Model OT-2 System uses
electrolytes which may contain substances that are
harmful if touched, swallowed, or inhaled. Avoid
contact with ANY fluid or powder in or around the
unit. What may appear to be plain water could
contain one of these toxic substances. In case of
eye contact, immediately flush eyes with water for
at least 15 minutes. Call physician. {See Appendix -
Material Safety Data Sheet (MSDS)}.
INSULATION
B-2C
SENSOR
CAP