Maintenance

Model 9060L

 

 

 

3.Insert a small flat-blade screwdriver into the slot in the receptacle wall nearest the fuse and gently pry open the fuse receptacle. The fuse holder will slide out. There are two fuses in use and are visible in the clip.

4.Remove the bad fuse and replace it with a 5x20mm 2A, 250 VAC, IEC time lag (T) fuse (P /N F1296).

5.Replace the fuse holder into its receptacle, pushing in firmly until it clicks.